- Robust Vacuum gauge for rough and fine vacuum range
- Rugged combination sensor combines most advanced Pirani and capacitance membrane technology
- High accuracy and excellent resolution in the range atmosphere to 5x10-5 mbar
- Standard or chemical resistant sensor version with built-in LCD Display
- Available as a package with all accessories for fine vacuum measurement
Multi-range vacuum gauge VMpro 1
DN16KF, 100-240V 50/60Hz
VMpro 1 multi-range vacuum gauge combines Pirani technology with the advantages of a ceramic capacitance diaphragm sensor in a single gauge.
In the measurement range between 10 mbar and atmospheric pressure the capacitance diaphragm technology provides gas-type independent, highly accurate values for reliable pressure measurement.
The built-in digital LCD display gives clear readability for easy vacuum level monitoring and also shows the current measuring principle. There is a status indication light on top of the sensor
Scope of delivery: VMpro 1 vacuum gauge, AC/DC power adapter with EU and UK plug adapters, connection cable
Multi-range vacuum gauge VMpro 2
Chemical resistant, DN16KF, 100-240V 50/60Hz
VMpro 2 chemical resistant multi-range vacuum gauge combines Pirani technology with the advantages of a ceramic capacitance diaphragm sensor in a single gauge.
In the measurement range between 10 mbar and atmospheric pressure the capacitance diaphragm technology provides gas-type independent, highly accurate values for reliable pressure measurement.
A Nickel Pirani filament is used to give good compatibility with a wide range of chemicals.
The built-in digital display gives clear readability for easy vacuum level monitoring and also shows the current measuring principle. There is a status indication light on top of the sensor.
Scope of delivery: VMpro 2 Chem vacuum gauge, AC/DC power adapter with EU and UK plug adapters, connection cable
Technical Data
Parameter | Data |
|---|---|
| VMpro 1 / VMpro 2 |
Measuring range [mbar] | 1500 – 5x10-5 |
Measuring range [Torr] | 1125 – 3.8x10-5 |
ACCURACY [% of reading] |
|
5x10-4 – 1x10-3 mbar | ±50 |
1x10-3 – 100 mbar | ±15 |
100 – 950 mbar | ±5 |
950 – 1050 mbar | ±2.5 |
Repeatability [% of reading] |
|
1x10-3 – 1100 mbar | ±2 |
Admissible temperature [OC] |
|
Operation | +10 – +50 |
Storage | -20 – +65 |
Bakeout at flange | ≤ +80 |
Bakeout at flange long tube | ≤ +250 |
Max Pressure [bar] | 10 |
Vacuum connection [DN] | 16 KF |
Maximum altitude [m] | 2000 |
Relative humidity, annual mean [%] (non-condensing) | ≤ 65 |
Voltage [V] | 100 - 240 |
Frequency [Hz] | 50 / 60 |
Plug adapters | EU (Schuko) & UK |
Mounting orientation | Any |
Internal volume [cm3] | 4.7 |
Sensor weight [g] | < 120 |
Sensor dimensions W/D/H [mm] (including cable bend) | 159 / 45 / 28 |